SU-D-217BCD-01: Corrupted DICOM Image Recovering: A Clinical Experience
L Chen1, A Walz-Flannigan1, S Langer1
1Mayo Clinic, Rochester, MN.
Purpose:
Colored DICOM secondary capture images generated from CT perfusion studies were corrupted if they were sent directly from a Siemens acquisition workstation to a GE viewing workstation. However, those images were properly displayed in the GE viewing workstation if they were transferred through a GE PACS first. The purpose of this work is to investigate the cause of image corruption and determine why passing through PACS corrected it.
Methods:
DICOM headers of corrupted and non-corrupted (sent through the PACS) images were compared with a free DICOM software tool (http://DVTK.org); the differences were highlighted. Certain header tags were found in non-corrupted images, but not in corrupted images. These tags were sequentially removed until the non- corrupted image became corrupted. Once a candidate tag was found, fresh corrupt images were modified by adding a 'repair' tag and tested.
Results:
It was found that the absence of Planar Configuration (0028, 0006) is the cause of image corruption. This attribute is used in the DICOM color image to specify whether the color pixel data are sent color-by-plane or color-by- pixel and should be present if the Sample per Pixel (0028, 0002) tag has a value greater than 1. In our DICOM color images, the values of (0028, 0002) and Photometric Interpretation (0028, 0004) are 3 and RGB, respectively. Thus (0028, 0006) should equal 0 (color-by-pixel), which is used for uncompressed or lossless compressed transfer syntaxes. Adding this tag and setting the value to zero manually repaired corrupt images.
Conclusions:
Using open source DICOM tools and following the described process can be a valuable ally in the search for causes of image corruption. Comparing the headers and finding the handful of different tags rapidly led to an explanation that could be used by the vendor for a permanent fix.
Related Concept Videos
Imaging Studies III: Computed Tomography
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...