SU-D-218-05: Material Quantification in Spectral X-Ray Imaging: Optimization and Validation
Related Experiment Video
Updated: Mar 2, 2026

Quantifying X-Ray Fluorescence Data Using MAPS
Published on: February 17, 2018
Purpose:
To develop and validate a multivariate statistical method to optimize scanning parameters for material quantification in spectral x-rayimaging.
Methods:
An optimization metric was constructed by extensively sampling the thickness space for the expected number of counts for m (two or three) materials. This resulted in an m-dimensional confidence region ofmaterial quantities, e.g. thicknesses. Minimization of the ellipsoidal confidence region leads to the optimization of energy bins. For the given spectrum, the minimum counts required for effective material separation can be determined by predicting the signal-to-noise ratio (SNR) of the quantification. A Monte Carlo (MC) simulation framework using BEAM was developed to validate the metric. Projection data of the m-materials was generated and material decomposition was performed for combinations of iodine, calcium and water by minimizing the z-score between the expected spectrum and binned measurements. The mean square error (MSE) and variance were calculated to measure the accuracy and precision of this approach, respectively. The minimum MSE corresponds to the optimal energy bins in the BEAM simulations. In the optimization metric, this is equivalent to the smallest confidence region. The SNR of the simulated images was also compared to the predictions from the metric.
Results:
TheMSE was dominated by the variance for the given material combinations,which demonstrates accurate material quantifications. The BEAMsimulations revealed that the optimization of energy bins was accurate to within 1keV. The SNRs predicted by the optimization metric yielded satisfactory agreement but were expectedly higher for the BEAM simulations due to the inclusion of scattered radiation.
Conclusions:
The validation showed that the multivariate statistical method provides accurate material quantification, correct location of optimal energy bins and adequateprediction of image SNR. The BEAM code system is suitable for generating spectral x- ray imaging simulations.
More Related Videos
10:12Synchrotron X-ray Microdiffraction and Fluorescence Imaging of Mineral and Rock Samples
Published on: June 19, 2018
10:42In Depth Analyses of LEDs by a Combination of X-ray Computed Tomography CT and Light Microscopy LM Correlated with Scanning Electron Microscopy SEM
Published on: June 16, 2016
Related Concept Videos
X-ray Imaging
X-ray Diffraction of Biological Samples
According to Bragg's law, when X-rays strike the sample positioned on a stage, the rays are scattered by the electron clouds around the sample atoms. The X-ray diffraction or scattering is caused by constructive interference of the X-ray waves that reflect off the internal...