SU-D-217BCD-03: Impact of Jaw Width on the Megavoltage Computed Tomography Image Quality and Imaging Dose with
Purpose:
On-treatment megavoltage computed tomography on Helical Tomotherapy (Accuray Inc., Sunnyvale, CA) is critical for image guided radiotherapy. A strategy was developed to assess the impact of various jaw widths on image quality and imaging dose with Tomotherapy.
Methods:
A cheese phantom (Gammex RMI, Middleton, WI) made of water equivalent materials was employed in this study. Three sets of measurements were independently carried out. Firstly, in the imaging dose measurement, the phantom was placed on the couch and aligned with a stationary green laser and beam isocenter. The measurement point was 10 mm up from the cente of the phantom. Three slices on either side of the middle slice were selected. Secondly, two inserts with different rows of holes of various sizes were placed inside the phantom for image contrast and resolution investigation. Lastly, twelve density inserts were placed into the outer holes in the phantom for measurement of the image value to density table (IVDT). A comparison of imaging dose, image resolution and contrast, IVDT table between different jaw configurations was performed to evaluate the imaging system.
Results:
Imaging dose was 2.93 cGy with a jaw size of one mm as opposed to 1.62 cGy with a four mm jaw, both of which are below the vendor's requirement: 3 cGy. However, image quality is improved significantly with the smaller jaw. Four lines of holes can be readily identified on images using smaller jaw while only three lines visible with the larger jaw. Image contrast is similarly enhanced when reducing the jaw size. On average CT numbers are 6% higher with the smaller jaw than those obtained with the larger one.
Conclusions:
Significant improvement in image quality is achieved with the smaller jaw field in Tomotherapy while the imaging dose is kept at a clinically acceptable level.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...