SU-D-217BCD-02: Online Angular Tube Current Modulation in X-Ray Computed Tomography: Can Tube Current Be Modulated
K Matsubara1,2, H Koshida1,2, T Takata1,2
1Kanazawa University, Kanazawa.
Purpose:
In X-ray computed tomography (CT), X-rays are significantly less attenuated in the anteroposterior direction and more in the lateral direction. Therefore, the tube current should be adjusted within one gantry rotation using angular tube current modulation (TCM). The aim of this study was to evaluate whether online angular TCM could reduce radiation dose appropriately.
Methods:
A 128-detector dual-source CT (SOMATOM Definition Flash; Siemens Healthcare, Erlangen, Germany) and an online TCM system (CARE Dose 4D; Siemens Healthcare) were used. Dose profiles were acquired using the CT Dose Profiler (RTI Electronics, Molndal, Sweden) and an elliptical cylindrical phantom (MHT; Kyoto Kagaku, Kyoto, Japan) for helical CT scans with and without TCM. In addition, absorbed dose distributions within a single section were acquired using an anthropomorphic phantom (RANI 10; The Phantom Laboratory, Salem, NY) and radiophotoluminescent glass dosimeters (RPLDs) (GD- 302M; Chiyoda Technol, Tokyo, Japan) for helical CT scans with and without TCM after placing RPLDs within all holes of one section and pasting them around the section. A graph of each absorbed dose distribution was drawn using graphing software (ORIGIN 8.6; OriginLab, Northampton, MA).
Results:
The acquired dose profiles suggested that online angular TCM could adjust the tube current in near-real time according to the attenuation measured from the previous projection. The profiles gradually stabilized because the tube current was adjusted properly. The absorbed doses were low and the distributions were stable with TCM compared with those without TCM.
Conclusions:
In X-ray CT, an online angular TCM can reduce radiation dose effectively by adjusting tube current within one gantry rotation in near-real time.
More Related Videos
10:42In Depth Analyses of LEDs by a Combination of X-ray Computed Tomography CT and Light Microscopy LM Correlated with Scanning Electron Microscopy SEM
Published on: June 16, 2016
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT