SU-E-T-467: Monte Carlo Dosimetric Study of the New Flexisource Co-60 High Dose Rate Source
J Vijande1,2,3, D Granero1,2,3, J Perez-Calatayud1,2,3
1University of Valencia, Valencia, Spain.
Purpose:
Recently, a new HDR 60Co brachytherapy source, Flexisource Co-60, has been developed (Nucletron B.V.). This study aims to obtain quality dosimetric data for this source for its use in clinical practice as required by AAPM and ESTRO.
Methods:
Penelope2008 and GEANT4 Monte Carlo codes were used to dosimetrically characterize this source. Water composition and mass density was that recommended by AAPM. Due to the high energy of the 60Co, dose for small distances cannot be approximated by collisional kerma. Therefore, we have considered absorbed dose to water for r<0.75 cm and collisional kerma from 0.75
Results:
TG-43 dosimetry parameters with L = 0.35 cm were obtained. Results performed with both radiation transport codes showed agreement typically within 0.2% for r > 0.8 cm and up to 2% closer to the source. Using Penelope2008 and GEANT4, an average of Î> = 1.085±0.003 cGy/(h U) (with k = 1, Type A uncertainties) was obtained. Dose rate constant, radial dose function and anisotropy functions for the Flexisource Co-60 are compared with published data for other Co-60 sources.
Conclusions:
Dosimetric data are provided for the new Flexisource Co-60 source not studied previously in the literature. Using the data provided by this study in the treatment planning systems, it can be used in clinical practice. This project has been funded by Nucletron BV.
More Related Videos
Related Concept Videos
Dose Size and Dosing Frequency: Determination Methods
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiation: Applications
The average...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...