Related Experiment Video
Updated: Mar 2, 2026

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
SU-E-J-60: Comparison of CT-On-Rails and a 3D Surface Imaging System for Image Guided Pelvic Radiation Therapy for
H Zhao1,2, B Wang1,2, P Rassiah-Szegedi1,2
1University of Utah, Salt Lake City, UT.
Purpose:
We investigate the feasibility of using AlignRT for pelvic radiation image guidance. The uniqueness of our study is that all patients have multiple CT-on-rails (CTOR) scans to compare corresponding AlignRT images to.
Methods:
Ten patients receiving pelvic radiation were enrolled in this study. Two simulation CT scans were performed in supine and prone positions for each patient. Body surface contours were generated in treatment planning system and exported to AlignRT to serve as reference images. The patient was aligned to treatment isocenter with room lasers, and then scanned with both CTOR and AlignRT in both supine and prone positions. Image guidance shifts were calculated for both modalities by comparison to the simulation CT and the differences between them were analyzed. These procedures were performed for each patient once per week for five weeks. The average and maximum difference of displacement between AlignRT and CTOR were calculated for each patient.
Results:
For supine position, there are 4 patients who had the average difference of displacement between AlignRT and CTOR along any direction (vertical, longitudinal, and lateral) greater than 0.5cm, and 1 patient greater than 1cm. For prone position, there are 7 patients who had the average difference greater than 0.5cm, and 3 patients greater than 1cm. For supine position, there are 4 patients who had the maximum difference great than 1cm. For prone position, there are 9 patients who had the maximum difference greater than 1 cm. The difference of displacement between AlignRT and CTOR is greater for prone position than for supine position.
Conclusions:
AlignRT does not appear to be an advisable image guidance approach for pelvic radiation therapy for patients with either supine or prone position. There appears to be a potential for large alignment discrepancies (up to 2.25 cm) between AlignRT and CTOR.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET