SU-E-J-10: Accuracy of Image Guidance Using Cone-Beam CT with Multiple Breath Hold Using Active Breathing
T Zhuang1, A Magnelli1, H Liu1
1The Cleveland Clinic Foundation, Cleveland, OH.
Purpose:
To complete a CBCT for a treatment using ABC, multiple breath hold (BH) (>3) were used due to the slow gantry rotation and the short BH period. Inter-BH tumor position variability may introduce distortion in the reconstructed images. This study aims to determine a threshold of the inter-BH scan displacement so that the inconsistency can be identified from the CBCT images.
Methods:
A numerical phantom was constructed to represent the thorax region of a human body. To simulate the inter-BH displacements, known magnitudes of motion (s = 0, 1, 3, 5 mm) along the longitudinal direction were introduced for the 'tumor' and 'diaphragm' in the phantom. Two different irregular motion patterns (s1=s3=/=s2 and s1=/=s2=/=s3) during CBCT scans were tested. Furthermore, a physical phantom with a movable insert was scanned using a commercial CBCT system. The insert of the phantom was programmed to move in the longitudinal direction according to the same motion patterns as designed in the numerical simulations. Subsequently, nine CBCT's in 'half-fan' mode for the physical phantom were acquired with the insert in various positions. These CBCT images were then fused to the reference CT by aligning to either the body of the phantom or the 'tumor' inside the insert.
Results:
Based on numerical simulation, position variation >1mm can be observed from the reconstructed CBCT images. Based on acquired CBCTs of the physical phantom, position variations of >3mm or 5mm were observed, depending on the motion pattern during the data acquisition. Because of the use of half-fan mode, we observed the order of position displacements of the tumor during CBCT acquisition drastically affected the outcome of imaging registration.
Conclusions:
Using ABC device, the inter-BH variability during a CBCT acquisition affects accuracy of tumor localization. A patient individualized planning margin might be necessary to account for this effect.
More Related Videos
08:54Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies III: Computed Tomography