SU-E-J-41: Fluoroscopy Based Adaptive Setup Approach for Thoracic Cancer IGRT.
T Chen1,2,1,1,1,3, S Qin1,2,1,1,1,3, S Jabbour1,2,1,1,1,3
1Cancer Institute of New Jersey, New Brunswick, NJ.
Medical Physics
|May 19, 2017
Summary
This study introduces a machine learning method using fluoroscopy images to precisely measure respiratory motion variations. This technique quantifies patient setup deviations for improved thoracic cancer image-guided radiation therapy (IGRT).
Area of Science:
- Medical Imaging
- Machine Learning
- Radiation Oncology
Background:
- Accurate patient positioning is crucial in thoracic cancer image-guided radiation therapy (IGRT).
- Respiratory motion during treatment can lead to significant setup deviations, impacting treatment efficacy and normal tissue toxicity.
- Real-time adaptive adjustments require precise quantification of motion and positional variations.
Purpose of the Study:
- To develop a fluoroscopy imaging-based approach for quantifying respiratory motion magnitude and phase variations.
- To determine online patient setup deviations in thoracic cancer IGRT.
- To enable real-time adaptive patient position adjustments.
Main Methods:
- A numerical phantom with simulated respiratory motion (varying frequency, amplitude, offset, phase shift) was used.
- White noise (SNR=5) was added to simulate clinical image quality.
- A manifold-based machine learning technique constructed a respiratory motion model, and MAP solutions quantified phase and position shifts.
Main Results:
- The approach successfully detected variations in motion patterns between image sets.
- It demonstrated insensitivity to frequency changes and image noise up to SNR=5.
- The method effectively captured and quantified changes in motion amplitude, position shift, and phase shift.
Conclusions:
- An effective mathematical approach quantifies motion differences between pre-treatment and planning images using machine learning on fluoroscopy.
- Application during online patient setup allows separation and adjustment of positioning deviations from respiratory motion.
- This facilitates minimized normal tissue toxicity in gated IGRT.