SU-E-J-09: Performance Optimization of Thick, Segmented Scintillators for Radiotherapy Imaging
Y El-Mohri1, R Choroszucha1, L Antonuk1
1University of Michigan, Ann Arbor, MI.
Thick segmented scintillators improve megavoltage imaging, but misalignment can reduce resolution. Selective binning techniques with high-resolution arrays restore spatial resolution and detective quantum efficiency (DQE) for better performance.
Area of Science:
- Medical Imaging Physics
- Detector Technology
- Radiological Sciences
Background:
- Thick segmented scintillators offer potential for enhanced megavoltage active matrix, flat-panel imagers (MV AMFPIs).
- Prototypes using CsI(Tl) and BGO have shown over a factor of 20 improvement in detective quantum efficiency (DQE) at zero spatial frequency.
- However, element misalignment and mis-registration to the imager array can degrade spatial resolution and limit DQE gains at higher frequencies.
Purpose of the Study:
- To investigate methods for restoring spatial resolution and DQE in MV AMFPIs with thick segmented scintillators.
- To address the challenges posed by element-to-element misalignment and mis-registration.
- To evaluate the efficacy of high-resolution active matrix, flat-panel imager (AMFPI) arrays and specialized binning techniques.
Main Methods:
- Theoretical and empirical investigation of misalignment and mis-registration effects on imaging performance.
- Determination of modulation transfer function (MTF) and DQE.
- Imaging of a phantom in a cone-beam CT geometry using a prototype BGO segmented scintillator (120x60 elements, 50 μm septa, 1016 μm pitch) coupled to a higher resolution 127-μm-pitch AMFPI array under a 6 MV photon beam.
Main Results:
- Misalignment and mis-registration significantly degrade spatial resolution, reducing DQE at non-zero spatial frequencies.
- Standard 8x8 pixel binning can correct for mis-registration in well-aligned scintillators but is affected by misalignment.
- Selective binning, focusing on interior pixels of each scintillator element, improves resolution while maintaining DQE.
Conclusions:
- High-resolution AMFPI arrays combined with selective binning enable prototype MV AMFPIs with thick segmented scintillators to achieve performance limited solely by scintillator characteristics.
- This approach overcomes resolution losses associated with misalignment and mis-registration.
- Further development can lead to improved diagnostic imaging capabilities in megavoltage applications.