SU-E-I-95: Multiscale Normalized Metal Artifact Reduction (MS-NMAR) in Computed Tomography
Purpose:
Medical x-ray CT devices produce images of high quality; however, the metal artifacts may arise from the adverse effect of metal materials present in the imaged objects. Conventional methods for metal artifact reduction (MAR) substitute the contaminated projection data corresponding to metal traces by specific interpolations in the projection space. Nevertheless, these methods usually introduce new artifacts, because the prior information of the imaged object (i.e., the prior image) was not involved. Therefore, a normalized MAR (NMAR) has further been invented, in which the projection space interpolation was carried out with respect to the ratio (i.e., normalization) of projection raw data to re-projection of the prior image. This feature renders NMAR a very high accuracy. However, the normalization in NMAR is more reasonable for the larger scale contents rather than for the smaller scale details.
Methods:
The proposed MS-NMAR method is a generalization of NMAR with a dedicated multiscale framework. Both projection raw data and re-projection of the prior image ill be decomposed into each scale. The normalization followed by the linear interpolation is performed in each scale, in which the larger the scale is the wider the interval the linear interpolation is operated with. Composite projection data are acquired by summing the projection components in all scales. MS-NMAR corrected image is reconstructed from the composite projection data.
Results:
Real CT data are used to verify the efficiency of the proposed method, and to compare with conventional MAR and NMAR. The quality of reconstructed image after MAR is evaluated by inspecting the region around the metal material.
Conclusions:
Both NMAR and MS-NMAR have a better performance as compared with the conventional MAR. Meanwhile, MS-NMAR outperforms NMAR, considering it hinders the productions of new artifacts while reducing the original metal artifacts. This work is partially supported by the US National Institute of Health through grants P50-AG025688 and 2P50AG025688.
More Related Videos
10:24Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021
07:13Author Spotlight: An Efficient and Robust Software for Automated Fusion of Multiple Preclinical Imaging Modalities
Published on: October 27, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography