SU-E-T-153: Detector-Grade CVD Diamond for Radiotherapy Dosimetry
Medical Physics
|May 19, 2017
Summary
Synthetic diamond films show promise as x-ray detectors for radiotherapy dosimetry. These chemical vapor deposition (CVD) diamond detectors offer high sensitivity and a small sensitive volume, making them suitable for precise dose measurements in small fields.
Area of Science:
- Medical Physics
- Materials Science
- Radiotherapy Physics
Background:
- Accurate dosimetry is critical in radiotherapy for effective cancer treatment.
- Current detectors may have limitations in sensitivity or spatial resolution for small radiation fields.
- Synthetic diamond films offer unique properties for radiation detection applications.
Purpose of the Study:
- To evaluate commercially available synthetic diamond films produced via chemical vapor deposition (CVD) as x-ray detectors.
- To assess their suitability for radiotherapy dosimetry, particularly in small radiation fields.
Main Methods:
- A CVD diamond detector with a 0.4 mm³ sensitive volume was fabricated and encapsulated.
- Measurements were performed in a water phantom using a clinical linear accelerator (linac) with 6- and 10-MV photons.
- Key detector characteristics evaluated included leakage current, response dynamics, dose linearity, dose rate dependence, percent depth dose, and output factors, with comparisons to ion chambers and silicon diodes.
Main Results:
- Negligible leakage currents (∼1 pA) and high sensitivity (680 nC/Gy) were observed.
- Detector response showed fast rise/fall times (< 2 s) and linearity up to 5 Gy.
- Percent depth dose profiles closely matched ion chamber measurements, and output factors were accurate for field sizes 5-15 cm.
Conclusions:
- CVD synthetic diamond films demonstrate significant potential for small field dosimetry in radiotherapy.
- Their small sensitive volume and high sensitivity are advantageous for precise dose measurements.
- Further research is needed to optimize detector packaging and explore broader clinical applications.