SU-E-T-167: Comprehensive Evaluation of EPID Image Acquisition for Integrating and Temporal Dosimetry of Fixed-Gantry IMRT and ArcIMRT
Related Experiment Video
Updated: Aug 6, 2026

Stem Cell Transplantation Strategies for the Restoration of Cognitive Dysfunction Caused by Cranial Radiotherapy
Published on: October 18, 2011
Purpose:
To evaluate EPID for dosimetry applications of arc and static-gantry IMRT with sliding window (SW) and/or step-and-shoot (SS) deliveries Methods: IMRT beams (SW & SS) were designed that generate beam hold-offs and dose rate modulation due to MLC motion under 10 × 10 cm jaw. An arcIMRT beam was designed by adding gantry movement to the SW field. A 10 cm × 10 cm open beam was also used. Despite differences in delivered dose rates/pulse characteristics, the four beams should deliver the same total dose. For each beam, various MUs with 6 MV beam at 300MU/min were irradiated on EPID which operated in image acquisition of integration mode (IM), continuous scanning mode with synchronization (CMs) and without (CMn) to beam pulses. Acquired images were evaluated in repeatability, dose linearity, and reproducibility (reproduce open beam output in IM).
Results:
In IM, repeatability, dose linearity, and reproducibility were within 1% for all dose levels and beams. In CMs, they were within 1-2% if dose rate was maintained steady (1) for SW beam (needed a minimum 1.3 MU/cm MLC motion) and (2) arcIMRT beam (needed a minimum 1 MU/degree and 2.8 MU/cm MLC motion) and (3) if a minimum of 38-40 MU per shoot was used for SS beam. Nonlinearity was observed for fewer MUs. This is due to the response of EPID to pulse-length reduction for fixed-gantry therapy and pulse dropping for arc therapy. The latter produces in-planar non-uniformity making EPID unsuitable for temporal dosimetry of arcIMRT. Sacrifice in temporal resolution then became necessary such as multi-frames per image (eg. ∼1 sec/image). In CMn the results were similar to those of CMs. However, they showed artifacts, thus this mode was not preferred.
Conclusions:
We found conditions under which integrating and temporal EPID dosimetry can be used for IMRT and arcIMRT dose deliveries.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies III: Computed Tomography