Summary
The carbon fiber couch top significantly increases patient skin dose, especially at smaller field sizes and higher energies. Accurate simulation in treatment planning systems is crucial for precise skin dose calculations.
Area of Science:
- Medical Physics
- Radiation Oncology
- Dosimetry
Background:
- Patient skin dose is clinically significant when the treatment couch-top contacts the skin.
- Radiation beams passing through the couch can alter dose distribution in the build-up region.
Purpose of the Study:
- To investigate the effect of a Varian carbon fiber exact couch-top on shallow depth dose.
- To evaluate the couch's contribution to patient skin dose in radiation therapy.
Main Methods:
- Dose measurements in the build-up region using a calibrated PTW chamber for 6MV and 18MV photon beams.
- Simulations in the Eclipse treatment planning system with adjusted couch CT numbers to match measurements.
Main Results:
- Surface dose increased with decreasing field size, with dose increases of 35-53% (6MV) and 50-113% (18MV) for field sizes from 20x20cm to 2x2cm.
- At 0mm depth, the dose difference due to the couch top reached 300%.
- The couch's effect on build-up dose extended to 1cm (6MV) and 2cm (18MV); optimal CT number is -200 to -300 HU.
Conclusions:
- Couch simulation in Eclipse improves dosimetric accuracy for couch-contacted skin dose.
- Accurate dose calculation requires consideration of surface definition, heterogeneity, resolution, and algorithm.
Keywords:
Alternating current power transmissionCalibrationCarbonDosimetryField sizeMedical treatment planningPhotonsRadiation therapyRadiation treatmentMore Related Videos
08:54Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
15.1K
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
7.8K