SU-E-T-176: Improved Collimator Scattering Factor (Sc) Measurements for Small Fields Using Build-Up Caps in Robotic
Medical Physics
|May 19, 2017
Summary
Build-up caps significantly impact in-air measurements for small fields in CyberKnife radiosurgery. Careful selection is crucial for commissioning data, especially for fields under 10 mm.
Area of Science:
- Medical Physics
- Radiotherapy
- Radiation Dosimetry
Background:
- CyberKnife robotic radiosurgery utilizes Monte Carlo calculations for treatment planning.
- In-air measurements are essential for the Monte Carlo algorithm's accuracy.
- The impact of build-up caps on these measurements, particularly for small fields, requires investigation.
Purpose of the Study:
- To compare the effect of different build-up cap thicknesses on in-air measurements using a diode detector.
- To evaluate the impact of build-up caps on Sc factors for small fields in CyberKnife radiosurgery.
Main Methods:
- Two custom-made acrylic build-up caps (1.5 cm and 5 cm) were used with an SFD diode detector.
- Measurements were performed in air at 80 cm SAD using a Blue Phantom.
- Comparisons were made across 12 fixed cones ranging from 5 mm to 60 mm.
Main Results:
- Significant differences (up to 20%) in Sc factors were observed for small fields (5-7.5 mm) with build-up caps compared to free-in-air measurements.
- Agreement within 2% was not achieved until cone sizes exceeded 40 mm.
- Observed differences were attributed to electron contamination, with both caps yielding similar Sc factors.
Conclusions:
- Build-up caps significantly affect Sc factors in small fields, a factor not explicitly addressed in CyberKnife planning manuals.
- The choice of build-up caps is critical for accurate commissioning data, especially for fields smaller than 10 mm.
- Further research into build-up cap selection for small field dosimetry is warranted.
Keywords:
CollimatorsMedical treatment planningMonte Carlo algorithmsRadiosurgeryRoboticsScattering measurementsSelected area electron diffractionMore Related Videos