SU-E-T-553: Dose-Mass Vs. Dose-Volume Optimization: A Phantom Study
I Mihaylov1,2,3, E Moros1,2,3, J Siebers1,2,3
1Rhode Island Hospital/Brown Medical Center, Providence, RI.
Mass-based optimization (DMH) is more general than volume-based optimization (DVH) in heterogeneous media. DMH accounts for density variations, improving dose delivery accuracy in radiotherapy planning.
Area of Science:
- Medical Physics
- Radiotherapy Planning
- Radiation Dosimetry
Background:
- Current radiotherapy planning often relies on dose-volume-histogram (DVH) based optimization.
- DVH optimization assumes uniform density, which can lead to inaccuracies in heterogeneous tissues.
Purpose of the Study:
- To compare the merits of mass-based optimization against volume-based optimization.
- To demonstrate the advantages of dose-mass-histogram (DMH) optimization in heterogeneous media using a simple phantom.
Main Methods:
- DVH-based objective functions were converted to DMH-based functions by incorporating voxel densities.
- A digital phantom with varying densities (0.2 g/cm³ and 0.8 g/cm³) was used.
- Monitor units were calculated for DVH- and DMH-based optimizations to achieve a target dose.
Main Results:
- Irradiating through lower density material (0.2 g/cm³) resulted in approximately 23% lower dose compared to higher density material (0.8 g/cm³).
- When DVH and DMH optimizations matched target dose-volume histograms, DMH delivered a higher proportion of dose through the lower density region (70% vs. 60%).
Conclusions:
- DMH optimization is more general and accurate than DVH optimization in heterogeneous media.
- DMH accounts for density variations, leading to improved dose deposition and reduced dose to critical structures.
- Mass-based optimization offers a more physically accurate approach to radiotherapy planning.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Related Concept Videos
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
Dosage Regimens: Partial Pharmacokinetic Parameters
Pharmacokinetic–Pharmacodynamic Relationship: Intensity of Dose-Effect Relationship
One-Compartment Open Model for IV Bolus Administration: Estimation of Elimination Rate Constant, Half-Life and Volume of Distribution