SU-E-T-528: Appraisal of Acorus XB and Convolution Dose Algorithms in Field Junction of Breast
Purpose:
Dose accuracy injunction regions of breast-tangential therapy is a challenge, and inaccurate dose predictions may lead to unreal hot/cold spots. Availability of the novel deterministic radiation transport method Acuros XB (AXB) provides a potential for more accurate dose predictions. This study assesses relative dose accuracies of this and the widely used other algorithms: collapsed cone convolution (CCC) and anisotropic analytical algorithm (AAA) against film measurements.
Methods:
A typical tangential and superclav fileld combination was planned for an anthropomorphic body phantom using Pinnacle-9.0 treatment-planning system (TPS). The created plan employing 6MV beam was delivered to the phantom on a Varian linac. In region of the field junction of tangentials & superclav, films (EBT2) were placed in coronal planes at two depths (∼ 2 and 4 cm). Optical density was measured along and ± 5mm away from the field-match line, and converted to dose using film-calibration curve specific to the batch of film. The same plan was also imported to Eclipse TPS using an import filter written in MATLAB. Algorithms Pinnacle CCC 9.0, Eclipse AAA 10.0.24 and AXB 11.0.3 were used for calculations. Comparison of the measured doses (assumed as gold standard) against doses calculated from planning-systems were preformed in a MATLAB platform.
Results:
In general, dose distributions from all three TPS algorithms are found to agree closely with film data. Agreements between AXB and CCC dose calculations were found to be reasonable. AXB appears to be better in modeling the backscatter effects in the heterogeneous regions. AAA calculations gives acceptable results, but with less accuracy compared to CCC and AXB.
Conclusions:
The novel deterministic algorithm AXB in Eclipse is found to provide better agreement with measured data in breast-tangential therapy. Benefits of using Acuors XB algorithm in tangential fields planning requires further investigation. This work was funded by National Institutes of Health through grant 2R44CA105806-02 and MD Anderson’s Cancer Center Support Grant CA0 16672.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies III: Computed Tomography
Imaging Studies II: Ultrasonography
Radiological Investigation I: X-ray and CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...