SU-E-T-516: Dosimetric Validation of AcurosXB Algorithm in Comparison with AAA & CCC Algorithms for VMAT Technique
M Kathirvel1, V Sai Subramanian1, G Arun1
1Yashoda Hospitals, Hyderabad, Andhra Pradesh, India.
The AcurosXB algorithm shows good agreement with AAA and CCC for Volumetric Modulated Arc Therapy (VMAT) dose calculations. This validated algorithm offers an accurate and efficient alternative for clinical use.
Area of Science:
- Medical Physics
- Radiation Oncology
- Computational Dosimetry
Background:
- Volumetric Modulated Arc Therapy (VMAT) requires accurate dose calculation algorithms for effective cancer treatment.
- Standard algorithms like Anisotropic Analytic Algorithm (AAA) and Collapsed Cone Convolution (CCC) are widely used but have limitations.
- The AcurosXB algorithm, a grid-based Boltzmann solver, offers potential for improved accuracy and speed.
Purpose of the Study:
- To dosimetrically validate the AcurosXB dose calculation algorithm for VMAT.
- To compare AcurosXB performance against the established AAA and CCC algorithms.
- To assess the accuracy and efficiency of AcurosXB in diverse clinical scenarios.
Main Methods:
- Five clinical VMAT cases each for Brain, Head&Neck, Thoracic, and Pelvic sites, plus SBRT, were analyzed.
- Dose predictions were performed using AcurosXB, AAA, and CCC algorithms.
- Measured doses in a multicube phantom with an iMatrixx-2D detector array served as the reference for comparison.
- Gamma analysis (3mm/3% and 2mm/2% criteria) was conducted to evaluate dose agreement.
Main Results:
- AcurosXB demonstrated strong agreement with AAA and CCC, with mean correlation coefficients around 0.998.
- Mean area gamma values for AcurosXB were 98.08% (3mm/3%) and 92.36% (2mm/2%), indicating high spatial accuracy.
- Mean average gamma values for AcurosXB were 0.28 (3mm/3%) and 0.42 (2mm/2%), signifying excellent dose agreement.
Conclusions:
- The AcurosXB algorithm provides dosimetrically accurate dose predictions for VMAT across various clinical sites.
- AcurosXB shows comparable and favorable agreement with AAA and CCC algorithms.
- AcurosXB is a valid, accurate, and efficient alternative for VMAT dose calculations in busy clinical settings.