TU-E-217BCD-07: Pilot Study on Consistency in Size Metrics for a Multimodality PEM/MR Breast Imaging Approach
M Sadinski1,1,1,2,2,1, M Giger1,1,1,2,2,1, K Drukker1,1,1,2,2,1
1University Chicago, Chicago, IL.
Purpose:
In this study, we evaluated the degree of consistency between size metrics obtained from PEM and MRI to determine the intrinsic effectiveness of a multimodality approach using these two systems for breast imaging.
Methods:
Under an IRB-approved protocol, 42 cases were considered (16 patients with 28 lesions), each consisting of an MRI data set and corresponding PEM data set, with an inclusion criterion of being obtained within 2 weeks of each other. Lesions were delineated manually on the PEM images and semi-automatically on the MRI images for efficiency. In addition to volume, equivalent sphere diameter (ESD) was evaluated for each lesion. This metric describes the diameter of a sphere with the same volume as that of the lesion, and is useful for minimizing the cubic impact of a single voxel contribution inherent to volume calculations.
Results:
The relationship between PEM based volumes and MRI based volumes showed a linear trend around VolMRI=VolPEM demonstrating a degree of consistency in the two volumes. The correlation between VolMRI and VolPEM was calculated as 0.547 with a corresponding p-value of 0.00018 demonstrating a significant correlation. The associated concordance was calculated as 0.534. The ESD metric showed a more significant linear trend with correlation 0.733 and corresponding p-value 3.5×10-8, indicating consistency in lesion size. The concordance is 0.717, indicating high reproducibility.
Conclusion:
Multimodality PEM/MRI breast imaging has the potential to combine functional and molecular imaging information for a powerful tool in cancer staging and evaluation of response to therapy. The resulting, expanded data set is of use only if fundamental size metrics are consistent between the two modalities. Our pilot data demonstrates that size metrics as we extracted from the image data are consistent between PEM and MRI breast image sets.Maryellen Giger is a stockholder in R2 Technology/Hologic, has equity in Quantitative Insights, and receives royalties from Hologic, GE Medical Systems, MEDIAN Technologies, Riverain Medical, Mitsubishi and Toshiba. It is the University of Chicago Conflict of Interest Policy that investigators disclose publicly actual or potential significant financial interest that would reasonably appear to be directly and significantly affected by the research activities.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
15:48Tracking the Mammary Architectural Features and Detecting Breast Cancer with Magnetic Resonance Diffusion Tensor Imaging
Published on: December 15, 2014
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging