SU-E-J-193: Gamma Knife Perfexion End-To-End Testing with a Commercial Stereotactic Phantom
Purpose:
To investigate the use of the Lucy ® Stereotactic Phantom (Standard Imaging, Inc.) for Gamma Knife Perfexion radiosurgery quality assurance of the imaging, treatment planning, and dose delivery processes. End-to-end testing of the Perfexion and Gamma Plan version 10.1 has not been previously examined in literature.
Methods:
The phantom was imaged using both the CT and T1- and T2-weighted MR sequences used for treatment planning. For imaging, the isocentric volume insert and fiducial markers were positioned within the phantom. Scans were transferred to the Gamma Plan treatment planning system and were evaluated for geometric and fusion accuracy. A plan was created to deliver 12Gy to the 50% isodose line to the 5.25cm3 volume. During dose delivery, Gafchromic EBT2 film was positioned in the film insert to replicate the position of the target volume. Dose results were analyzed using RIT software (Radiologic Imaging Technology, Inc.).
Results:
Image fusion integrity was inspected by overlaying the MR and CT markers (5 fiducial markers spaced 5mm apart) and visually examining the resulting volume insert overlap between the three scans. Geometric accuracy was evaluated by contouring three volumes using Gamma Plan contouring tools. Agreement within 1.1%, 6.7% and 12.2% of the actual volumes was seen with the T1-weighted, T2-weighted, and CT images, respectively. The volume-based acquisition and 1mm slice thickness of the T1-weighted sequence resulted in the most accurate measurement. Geometric measurements along two dimensions showed acceptable accuracy for all imaging modalities within 1.6%. Dosimetry results agreed well with the planned dose. The EBT2 film was calibrated for absolute dose measurements using a dose calibration curve for 0.1-30 Gy and the calibration curve was verified to have <3% error above 1Gy.
Conclusions:
The Lucy phantom allows for comprehensive quality assurance testing of the Gamma Knife Perfexion radiosurgery process.
More Related Videos
08:59Use of a Multi-compartment Dynamic Single Enzyme Phantom for Studies of Hyperpolarized Magnetic Resonance Agents
Published on: April 15, 2016
08:41Patient-Specific Polyvinyl Alcohol Phantom Fabrication with Ultrasound and X-Ray Contrast for Brain Tumor Surgery Planning
Published on: July 14, 2020
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):