SU-E-I-32: Sterile Radiation Reduction Gloves May Be Contraindicated in Fluoroscopically Guided Interventions
A Pasciak1,2,3, M Neveu1,2,3, A Sriharan1,2,3
1University of Tennessee Medical Center, Knoxville, TN.
Purpose:
Sterile radiation reduction gloves have been widely used in the past decade to provide modest decreases in operator hand dose when the hands are placed in the field of view (FOV). While multiple publications have quantified the potential dose reduction from the use of such gloves, possible effects on the patient have not yet been assessed. The aim of this study was to determine if radiation reduction gloves can Result in a significant increase in patient dose and increased risk of radiation induced skin injury when used in interventional radiology.
Methods:
The effect of radiation reduction gloves when used in the FOV was determined by measurement of patient entrance exposure rate (EER) for a variety of patient sizes and varying operating and magnification modes. EERs were measured with no glove in the FOV, with one glove and, to replicate the actions of many dose-conscious radiologists, with double gloves in the FOV.
Results:
Compared to an ungloved hand, the use of a single radiation reduction glove near the center of the FOV results in a 2-fold average increase in patient EER. The use of double radiation reduction gloves results in a 3-fold average increase in EER. In both cases, this increase was only weakly dependent on the size of the patient and on the operating and magnification modes used. In fact, patient thicknesses ranging from 6-14 inches and operating modes ranging from low-dose fluoroscopy to DSA produced less than a 20% deviation from the increases in EER quoted above.
Conclusions:
When used in the FOV, radiation reduction gloves can substantially increase patient EER. This increase in patient dose, when compared with the relatively small published reduction in extremity dose provided to the operator, may make their use contraindicated in cases where radiation induced skin injury is a possible risk.
More Related Videos
08:17Low-Dose Gamma Radiation Sterilization for Decellularized Tracheal Grafts
Published on: April 14, 2023
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Hand hygiene
Hand washing...
Radiological Investigation I: X-ray and CT
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...