SU-E-I-42: Measurement of X-Ray Beam Width and Geometric Efficiency in MDCT Using Radiochromic Films
T Liillau1,2, M Liebmann1,2, H von Boetticher1,2
1Medical Radiation Physics, Carl von Ossietzky University & Pius-Hospital, Oldenburg, Germany.
Medical Physics
|May 19, 2017
Summary
Radiochromic films accurately measured multi-detector computed tomography (MDCT) x-ray beam width and geometric efficiency (GE). Results show GE and over-beaming factors vary significantly with beam collimation, crucial for accurate CT dose calculations.
Area of Science:
- Medical Physics
- Radiological Imaging Science
Background:
- Multi-detector computed tomography (MDCT) scanners utilize 'over-beaming,' where the primary x-ray beam exceeds nominal collimation.
- This over-beaming increases patient radiation dose compared to single detector CT.
- Precise measurement of primary beam width and geometric efficiency (GE) is vital for accurate CT dose calculations using Monte Carlo or analytical methods.
Purpose of the Study:
- To measure the x-ray beam width and geometric efficiency (GE) of an MDCT scanner for various beam collimations.
- To assess the utility of radiochromic films for these measurements.
- To determine over-beaming factors for different collimations.
Main Methods:
- Radiochromic films (Gafchromic XR-QA2) were used to measure axial dose profiles in air for six different beam collimations.
- Measurements were performed on a Siemens SOMATOM Sensation 64 Scanner at 120 kVp.
- Films were calibrated against a PTW semiflex ionization chamber; geometric efficiency and full-width-at-half-maximum (FWHM) were determined.
Main Results:
- Geometric efficiency (GE) ranged from 58% to 85% for MDCT beam collimations from 7.2 mm to 28.8 mm.
- Over-beaming factors varied from 1.11 to 1.43.
- For smaller collimations (1x5 mm and 1x10 mm), GE was 77% and 84%, with over-beaming factors close to 1.
Conclusions:
- Radiochromic films provide accurate measurements of x-ray beam width and geometric efficiency due to their high spatial resolution.
- Both geometric efficiency and the over-beaming factor are strongly dependent on the selected beam collimation.
- These findings are essential for improving the accuracy of CT dose calculations.