SU-E-I-08: KV XVI Cone Beam-CT Dose Measurement Using Gafchromic XRQA2 Film
T Giaddui1,2,3, Y Cui1,2,3, Z Yegingil1,2,3
1Thomas Jefferson University, Philadelphia, Pennsylvania.
Purpose:
To study the effect of different filters on the dose response curves of the Gafchromic XRQA2 film. To measure the kV XVI cone-beam CT (CBCT) surface dose received during 3D and 4D imaging protocols in three body regions (head and neck, chest and pelvis).
Methods:
GafChromic XR- QA2 film (International Specialty Products, Wayne, NJ) dose response curves were generated for three irradiation settings: 100 kVp S20/F0; 120 kVp S20/F0 and 120 kVp S20/F1(F1 is a Bowtie filter). Film pieces were irradiated in air by the X-ray Volume Imager (XVI) mounted on the Elekta Synergy linear accelerator (Elekta, Crawley, UK) and their responses were correlated to air kerma measurements. To measure the CBCT surface dose, film pieces were taped on the surface of a male Alderson Rando Phantom (Alderson Research Laboratories, Inc., Long Island City, New York) at four different places (Anterior, Posterior, Right Lateral, Left Lateral).
Results:
The dose response curves of XRQA2 film generated with F1 and F0 filters were found to differ by 5 to 7% when the air kerma changed between 2 and 5 cGy. This was less than the observed difference (more than 15%, especially at low air kerma) in the dose response curves when different energies (100 and 120 kVp) and same filter were used. Surface dose ranged between 0.02 cGy and 4.99 cGy. The lowest average surface dose (0.05 cGy) was observed when the fast head and neck protocol was used, whilst the highest average surface dose (3.06 cGy) was noticed when the chest m2 0 protocol was used.
Conclusions:
Filters seem to have less effect on the dose response of the film compared with energy. Gafchromic XRQA2 film was used successfully to measure the XVI CBCT surface dose. The dose was found to vary from one imaging protocol to another, with 4D protocols not necessarily delivering more doses.
More Related Videos
Related Concept Videos
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT