Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
WE-G-213CD-06: Implementation of Real-Time Tumor Tracking Using Robotic Couch
I Buzurovic1,2, Y Yu1,2, T Podder1,2
1Thomas Jefferson University, Philadelphia, PA.
This study presents a novel robotic couch system for real-time tumor tracking, achieving sub-millimeter accuracy. This method enhances radiation therapy by minimizing damage to healthy tissues.
Area of Science:
- Medical Physics
- Robotics in Medicine
- Radiation Oncology
Background:
- Accurate patient positioning is crucial in radiation therapy.
- Existing robotic couches lack real-time tumor motion compensation.
- Tumor motion during treatment can lead to geographic misses and increased toxicity.
Purpose of the Study:
- To introduce a novel method for real-time tumor tracking using a modified commercial robotic treatment couch.
- To evaluate the accuracy and feasibility of this tumor tracking system.
Main Methods:
- A commercially available robotic couch (Elekta PreciseTM Table) was adapted with a novel control system.
- The couch was equipped with Maxon and Rockwell Automation motors and Baumer encoders for 3D motion control.
- Galil controllers and Advance Motion Controls amplifiers were integrated for precise position and velocity adjustments.
Main Results:
- The system demonstrated real-time tracking capabilities within a 50mm range in all three directions (superior-inferior, lateral, anterior-posterior).
- Achieved tracking accuracies were 0.15mm (superior-inferior), 0.20mm (lateral), and 0.18mm (anterior-posterior).
- The system exhibited motion repeatability within ±0.2mm.
Conclusions:
- The developed system enables real-time tumor tracking with high, sub-millimeter accuracy.
- This technique offers a potentially simple and effective approach to reduce irradiation of healthy tissues.
- The findings support the integration of advanced robotic couch systems for improved radiation therapy precision.