Four-Dimensional CT Analysis Using Sequential 3D-3D Registration
Published on: November 23, 2019
SU-E-J-87: Building Deformation Error Histogram and Quality Assurance of Deformable Image Registration
1Case Western Reserve University, Radiation Oncology Department, Cleveland, OH.
Purpose:
To quantify the systematic error of a Deformable Image Registration (DIR) system and establish Quality Assurance (QA) procedure.
Methods:
To address the shortfall of landmark approach which it is only available at the significant visible feature points, we adapted a Deformation Vector Map (DVM) comparison approach. We used two CT image sets (R and T image sets) taken for the same patient at different time and generated a DVM, which includes the DIR systematic error. The DVM was calculated using fine-tuned B-Spline DIR and L-BFGS optimizer. By utilizing this DVM we generated R' image set to eliminate the systematic error in DVM,. Thus, we have truth data set, R' and T image sets, and the truth DVM. To test a DIR system, we use R' and T image sets to a DIR system. We compare the test DVM to the truth DVM. If there is no systematic error, they should be identical. We built Deformation Error Histogram (DEH) for quantitative analysis. The test registration was performed with an in-house B-Spline DIR system using a stochastic gradient descent optimizer. Our example data set was generated with a head and neck patient case. We also tested CT to CBCT deformable registration.
Results:
We found skin regions which interface with the air has relatively larger errors. Also mobile joints such as shoulders had larger errors. Average error for ROIs were as follows; CTV: 0.4mm, Brain stem: 1.4mm, Shoulders: 1.6mm, and Normal tissues: 0.7mm.
Conclusions:
We succeeded to build DEH approach to quantify the DVM uncertainty. Our data sets are available for testing other systems in our web page. Utilizing DEH, users can decide how much systematic error they would accept. DEH and our data can be a tool for an AAPM task group to compose a DIR system QA guideline. This project is partially supported by the Agency for Healthcare Research and Quality (AHRQ) grant 1R18HS017424-01A2.
More Related Videos
08:39Longitudinal Micro-Computed Tomography Image Analysis for User-Defined Region of Interest in Critical-Sized Bone Defects
Published on: June 24, 2025
09:34A Method to Estimate Cadaveric Femur Cortical Strains During Fracture Testing Using Digital Image Correlation
Published on: September 14, 2017
Related Concept Videos
Deformation of Member under Multiple Loadings
In the case of a member with a variable cross-section, the strain is not constant but depends on the position. The deformation of an...
Deformation of a Beam under Transverse Loading
The insights from the bending moment diagram extend to...