Development and validation of a highly sensitive gradient chiral separation of pramipexole in human plasma by
Marco Michi1, Massimo Breda1, Justin Keane Belardi2
1Aptuit Verona Srl, Via Fleming 4, 37135 Verona, Italy.
Aim:
Development of a high-sensitivity chiral LC-MS/MS method was required to evaluate a combination of pramipexole (S-PPX) and its enantiomer dexpramipexole (R-PPX) in a proposed clinical trial. The previously available methods suffered from low sensitivity for the (S)-enantiomer in the presence of the more abundant (R)-enantiomer. Based on the projected dosing regimen in the clinical trial, a 5000-fold improvement in sensitivity was required for the (S)-enantiomer.
Methodology:
Spiked human plasma samples were extracted by liquid-liquid extraction using ethyl acetate and injected onto a CHIRALPAK ID column under pH gradient conditions.
Conclusion:
An improved analytical method was developed and validated with a final LLQ for (S)-PPX of 0.1 ng/ml in the presence of 2000 ng/ml of (R)-PPX.
More Related Videos
07:08A High Throughput, Multiplexed and Targeted Proteomic CSF Assay to Quantify Neurodegenerative Biomarkers and Apolipoprotein E Isoforms Status
Published on: October 20, 2016
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
Related Concept Videos
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Types of Detectors
High-Performance Liquid Chromatography: Instrumentation