Magnetic Resonance Derived Myocardial Strain Assessment Using Feature Tracking
Published on: February 12, 2011
Intervendor Differences in the Accuracy of Detecting Regional Functional Abnormalities: A Report From the EACVI-ASE
Oana Mirea1, Efstathios D Pagourelias1, Jurgen Duchenne1
1Department of Cardiovascular Diseases, University Hospital Leuven, Leuven, Belgium.
Vendor-specific strain analysis tools show varying accuracy in detecting myocardial abnormalities after infarction. Differences among vendors highlight the need for careful software selection in clinical practice for precise regional function assessment.
Area of Science:
- Cardiology
- Medical Imaging
- Echocardiography
Background:
- Speckle tracking echocardiography (STE) is a promising tool for quantitative assessment of regional myocardial function.
- Limited research exists on the accuracy differences among STE software for identifying regional abnormalities.
Purpose of the Study:
- To compare the accuracy of vendor-specific and independent strain analysis tools in detecting regional myocardial function abnormalities.
- To evaluate the performance of different STE software in a clinical setting.
Main Methods:
- Sixty-three subjects (5 healthy, 58 patients with myocardial infarction) were examined using 7 ultrasound machines.
- Segmental strain parameters (peak systolic, end-systolic, post-systolic) were measured using 6 vendor-specific and 2 independent software tools.
- Strain parameters were compared between scarred and scar-free segments; receiver-operating characteristic curves assessed discrimination ability.
Main Results:
- All vendors detected significantly lower strain values in scarred segments compared to normal segments (p < 0.05).
- Accuracy in identifying transmural scar ranged from acceptable to excellent (AUC 0.70-0.83).
- Significant differences in accuracy were found among vendors (p < 0.05), with lower accuracy for basal segments compared to apical segments.
Conclusions:
- The accuracy of identifying regional myocardial abnormalities significantly differs among STE software vendors.
- Clinical application of STE requires careful consideration of software-specific performance variations.