Related Experiment Video
Updated: Mar 2, 2026

Radiotracer Administration for High Temporal Resolution Positron Emission Tomography of the Human Brain: Application to FDG-fPET
Published on: October 22, 2019
Baseline and longitudinal variability of normal tissue uptake values of [18F]-fluorothymidine-PET images
Matthijs C F Cysouw1, Gerbrand M Kramer1, Virginie Frings1
1Department of Radiology & Nuclear Medicine, VU University Medical Center, Amsterdam, The Netherlands.
Purpose:
[18F]-fluorothymidine ([18F]-FLT) is a PET-tracer enabling in-vivo visualization and quantification of tumor cell proliferation. For qualitative and quantitative analysis, adequate knowledge of normal tissue uptake is indispensable. This study aimed to quantitatively investigate baseline tracer uptake of blood pool, lung, liver and bone marrow and their precision, and to assess the longitudinal effect of systemic treatment on biodistribution.
Methods:
18F-FLT-PET(/CT) scans (dynamic or static) of 90 treatment-naïve oncological patients were retrospectively evaluated. Twenty-three patients received double baseline scans, and another 39 patients were also scanned early and late during systemic treatment with a tyrosine kinase inhibitor. Reproducible volume of interest were placed in blood pool, lung, liver, and bone marrow. For semi-quantitative analysis, SUVmean, SUVmax, and SUVpeak with several normalizations were derived.
Results:
SUVs of basal lung, liver, and bone marrow were not significantly different between averaged dynamic and static images, in contrast with blood pool and apical lung. Highest repeatability was seen for liver and bone marrow, with repeatability coefficients of 18.6% and 20.4% when using SUVpeak. Systemic treatment with TKIs both increased and decreased normal tissue tracer uptake at early and late time points during treatment.
Conclusion:
Simultaneous evaluation of liver and bone marrow uptake in longitudinal response studies may be used to assess image quality, where changes in uptake outside repeatability limits should trigger investigators to perform additional quality control on individual PET images.
Advances In Knowledge:
For [18F]-FLT PET images, liver and bone marrow have low intra-patient variability when quantified with SUVpeak, but may be affected by systemic treatment.
Implications For Patient Care:
In [18F]-FLT-PET response monitoring trials, liver and bone marrow uptake may be used for quality control of [18F]-FLT PET images.
More Related Videos
10:04Radionuclide-fluorescence Reporter Gene Imaging to Track Tumor Progression in Rodent Tumor Models
Published on: March 13, 2018
10:02Quantification of Atherosclerotic Plaque Activity and Vascular Inflammation using [18-F] Fluorodeoxyglucose Positron Emission Tomography/Computed Tomography FDG-PET/CT
Published on: May 2, 2012
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...