Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
EIT based pulsatile impedance monitoring during spontaneous breathing in cystic fibrosis
Sabine Krueger-Ziolek1,2, Benjamin Schullcke1,2, Bo Gong1,2
1Institute of Technical Medicine, Furtwangen University, Jakob-Kienzle-Straße 17, 78054 Villingen-Schwenningen, Germany.
Objective:
Evaluating the lung function in patients with obstructive lung disease by electrical impedance tomography (EIT) usually requires breathing maneuvers containing deep inspirations and forced expirations. Since these maneuvers strongly depend on the patient's co-operation and health status, normal tidal breathing was investigated in an attempt to develop continuous maneuver-free measurements.
Approach:
Ventilation related and pulsatile impedance changes were systematically analyzed during normal tidal breathing in 12 cystic fibrosis (CF) patients and 12 lung-healthy controls (HL). Tidal breaths were subdivided into three inspiratory (In1, In2, In3) and three expiratory (Ex1, Ex2, Ex3) sections of the same amplitude of global impedance change. Maximal changes of the ventilation and the pulsatile impedance signal occurring during these sections were determined (▵I V and ▵I P). Differences in ▵I V and ▵I P among sections were ascertained in relation to the first inspiratory section. In addition, ▵I V/▵I P was calculated for each section.
Main Results:
Medians of changes in ▵I V were <0.05% in all sections for both subject groups. Both groups showed a similar pattern of ▵I P changes during tidal breathing. Changes in ▵I P first decreased during inspiration (In2), then increased towards the end of inspiration (In3) and reached a maximum at the beginning of expiration (Ex1). During the last two sections of expiration (Ex2, Ex3) ▵I P changes decreased. The CF patients showed higher variations in ▵I P changes compared to the controls (CF: -426.5%, HL: -158.1%, coefficient of variation). Furthermore, ▵I V/▵I P significantly differed between expiratory sections for the CF patients (Ex1-Ex2, p < 0.01; Ex1-Ex3, p < 0.001; Ex2-Ex3, p < 0.05), but not for the controls. No significant differences in ▵I V/▵I P between inspiratory sections were determined for both groups.
Significance:
Differences in ▵I P changes and in ▵I V/▵I P between both subject groups were speculated to be caused by higher breathing efforts of the CF patients due to airway obstruction leading to higher intrathoracic pressures, and thus to greater changes in lung perfusion.
More Related Videos
05:56Implementation of Non-invasive Point of Care Transient Elastography for Evaluation of Liver Disease in Pediatric Populations with Cystic Fibrosis
Published on: August 29, 2025
11:13Collection, Expansion, and Differentiation of Primary Human Nasal Epithelial Cell Models for Quantification of Cilia Beat Frequency
Published on: November 10, 2021
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration...