Validation of Navigation Ultrasound for Clavicular Length Measurement
Anders Thorsmark Høj1, Chiara Villa2, Ole M Christensen3
1Department of Orthopedic Surgery and Traumatology, Holbæk Hospital, Holbæk, Denmark; Orthopedic Research Unit, Department of Orthopedic Surgery and Traumatology, Odense University Hospital, Odense, Denmark.
Abstract:
Accurate clavicular length measurement is considered important for the selection of patients requiring surgical treatment after a midclavicular fracture. As radiographic length measurements can be inaccurate to varying degrees, new measurement methods are needed. Use of novel technology, navigation ultrasound, to deliver accurate measurements seems promising. To examine the reliability and validity of navigation ultrasound for length measurements of the clavicle, a study was performed using length measurements performed before forensic autopsies from 2012 to 2013. In total, 38 bodies with a median age of 52.5 (range: 21-78 y) were included. Navigation ultrasound exhibited high reliability (intra-class correlation coefficient: 0.942-0.997, standard error of the mean: 0.7-2.9 mm, minimal detectable change: 2.3-8.1 mm) and validity (measurement error: 1.3%-1.8%, limits of agreement (95% confidence interval): approximately ± 7.5 mm, Pearson's correlation R: 0.948-0.974). Navigation ultrasound can measure clavicular length with an intra-rater reliability matching that of 3-D rendered computed tomography scans and with high validity. Its use could spread to other fields requiring accurate musculoskeletal length measurements.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: