Related Experiment Video
Updated: Mar 2, 2026

Author Spotlight: Microfluidic Channel-Based Soft Electrodes and Their Application in Capacitive Pressure Sensing
Published on: March 17, 2023
An Angstrom-sensitive, differential MEMS capacitor for monitoring the milliliter dynamics of fluids
David J Apigo1, Philip L Bartholomew2, Thomas Russell1
1New Jersey Institute of Technology, Department of Physics, Newark, NJ 07102, USA.
Abstract:
A device, with MEMS sensors at its core, has been fabricated and tested for measuring low fluid pressure and slow flow rates. The motivation was to measure clinically relevant ranges of slow-moving fluids in living systems, such as the cerebrospinal fluid in the brain. For potential clinical utility, the device can be read transcutaneously by inductive coupling to MEMS capacitive sensors in circuits with resonance frequencies in the MHz range. Signal shifts for flow rates in the range of 0-42 mL/h and differential pressure levels between 0.1 and 2 kPa have been measured, because the sensitivity in the capacitance gap measurement is about 1 Å. The sensors have been used successfully to monitor simulated cerebrospinal fluid dynamics. The device does not utilize any internal power, since it is powered externally via the inductive coupling.