Published on: September 29, 2020
Oral capnography is more effective than nasal capnography during sedative upper gastrointestinal endoscopy
Wei-Nung Teng1,2, Chien-Kun Ting1, Yu-Tzu Wang2
1Department of Anaesthesiology, Taipei Veterans General Hospital and National Yang-Ming University, No. 201, Sec 2, Shipei Rd., Beitou Dist., Taipei City, 11217, Taiwan, ROC.
Abstract:
The role of capnography in esophagogastroduodenoscopy (EGD) is controversial. Simultaneous supplemental oxygen, position of patient, open mouth breathing pattern, and anatomy of the oral and nasal cavity can influence capnography accuracy. This study first measured capnographic data via the nasal or oral cavity during sedated EGD. Secondly, we investigated the influence of supplementary oxygen through the oral cavity on the capnographic reading. Patients with ASA class I or II status admitted for routine EGD exams were enrolled. End-tidal carbon dioxide measurements were performed simultaneously via nasal catheter and oral catheter with standard oral bite and nasal cannula supplementary oxygen when the patient is awake, during sedation and during sedation with endoscopy. The influence of oral supplementary oxygen, oral capnography were recorded using a mandibular advancement bite block. One hundred and four patients were enrolled. Breathing in the conscious patient is conducted primarily via the nostrils (95%). When sedated with endoscope placement, the percentage of nasal breathing decreased significantly to 47% and oral capnography sufficiently captured data in 100% of patients. Supplementary oral oxygen decreased oral capnographic measurement significantly (38.89 ± 7.148 vs. 30.73 ± 7.84, p < 0.001). However, the measurements using the MA bite block did not differ from oral cavity catheter (28.86 ± 8.51 vs. 30.73 ± 7.839, p = 0.321). The conscious patient breathes mostly nasally while the sedated patient breathes mostly orally during EGD when an oral bite is in place. Capnography measurement via oral cannula increases the measurement accuracy and efficacy. Oral supplementary oxygen may decrease capnographic measurement but still provide sufficient reading for interpretation.
More Related Videos
07:52Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
07:54Drug-Induced Sleep Endoscopy DISE with Target Controlled Infusion TCI and Bispectral Analysis in Obstructive Sleep Apnea
Published on: December 6, 2016
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Endoscopic Procedures III: Video Capsule Endoscopy
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques