Accurate determination of brain metabolite concentrations using ERETIC as external reference
Niklaus Zoelch1,2,3, Andreas Hock1,3, Susanne Heinzer-Schweizer1
1Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland.
NMR in Biomedicine
|May 26, 2017
Summary
The Electric REference To access In vivo Concentrations (ERETIC) method accurately quantifies in vivo metabolite concentrations in the human brain. This improved ERETIC approach corrects for magnetic field inhomogeneities, offering advantages over traditional water referencing, especially in diseased tissues.
Area of Science:
- Neuroimaging
- Biophysics
- Medical Physics
Background:
- Magnetic Resonance Spectroscopy (MRS) enables in vivo metabolite quantification using a reference signal.
- Human brain 1H MRS commonly uses tissue water as a reference, but requires accurate water concentration estimation.
- ERETIC method offers an alternative by comparing metabolite signals to a phantom reference, correcting for coil loading.
Purpose of the Study:
- To enhance the accuracy of the ERETIC method by correcting for spatial B1 inhomogeneities.
- To simplify ERETIC quantification through automatic phase correction.
- To validate the improved ERETIC method for in vivo metabolite concentration determination.
Main Methods:
- Implemented B1 correction using volume-selective power optimization and contrast-minimized brain images.
- Corrected reception sensitivity variations by aligning phantom and in vivo voxel positions.
- Incorporated automatic phase correction for the ERETIC signal.
Main Results:
- Metabolite concentrations determined by the enhanced ERETIC method in 21 healthy subjects agreed with those derived using tissue water as a reference.
- Brain water concentrations measured with ERETIC correlated well with tissue segmentation and literature values.
- The B1 correction scheme improved the accuracy of ERETIC-based metabolite quantification.
Conclusions:
- The refined ERETIC method provides accurate in vivo metabolite concentrations without assumptions about tissue water content.
- This technique offers potential advantages over internal water referencing, particularly for studying diseased brain tissues.
- The enhanced ERETIC method is a validated tool for reliable in vivo metabolite quantification.