Multi-Scale Modification of Metallic Implants With Pore Gradients, Polyelectrolytes and Their Indirect Monitoring In vivo
Published on: July 1, 2013
Mechanical properties of a nanoporous membrane used in implantable medical devices. Correlation between experimental
F Cristofari1, B Piotrowski1, R Pesci1
1ENSAM-Arts et Métiers ParisTech, LEM3 UMR CNRS 7239, 4 rue Augustin Fresnel, 57078 Metz Cedex 03, France.
Abstract:
Nanoporous membranes are used for the elaboration of implantable medical devices. In order to guaranty their integrity after implantation in a patient body, it is necessary to characterize the microstructure and the mechanical behavior of such membranes. They present randomly distributed pores around 1µm in diameter at the surface. X-ray nanotomography permits to get the geometry of the pores through the thickness with a reduction of the diameter in the core. A multiscale study is done to characterize the membranes: macroscopic tensile tests permit to get the behavior law of the non porous material and in situ tensile tests are carried on in a Scanning Electron Microscope in order to observe the evolution of pores and cracks during loading. A 2D Finite Element Model is also developed in parallel. The confrontation between experiments and numerical simulations permit to validate the accuracy of the model. The latter is then used to simulate several types of loadings considering various pore distributions and sizes.