Clinical Implications of Different Blood Pressure Measurement Techniques
1Division of Renal Diseases & Hypertension, University of Minnesota, 717 Delaware St SE, Suite 353, Minneapolis, MN, 55414, USA. Draw0003@umn.edu.
Purpose Of Review:
The appropriate treatment targets for individuals with elevated blood pressure (BP) have received increased attention in light of recent clinical trial results. However, it is well known that the method used to measure BP can have a significant impact on the observed BP. In this review, we summarize the existing literature on the impact of BP measurement technique on observed BP readings.
Recent Findings:
Manual BPs obtained in-clinic routinely differ from those obtained using automated devices. Further, clinic-based readings (either manual or automated) typically correlate poorly with readings from ambulatory BP monitoring or home-based devices. However, few studies utilize randomization or sound experimental design to explore differences in BP readings by method or technique. While numerous studies report differences in BP by method, most lack statistical rigor and therefore provide limited insight into the true effect of technique on BP measurements.
More Related Videos
07:35Simultaneous Electrocardiography Recording and Invasive Blood Pressure Measurement in Rats
Published on: January 31, 2019
09:38Femoral Arterial and Venous Catheterization for Blood Sampling, Drug Administration and Conscious Blood Pressure and Heart Rate Measurements
Published on: January 24, 2012
Related Concept Videos
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)