Functional atrial undersensing associated with device algorithm promoting AV conduction
Related Experiment Video
Updated: Mar 1, 2026

Optimization of Transesophageal Atrial Pacing to Assess Atrial Fibrillation Susceptibility in Mice
Published on: June 29, 2022
Pasquale Crea1, Angela Nicotera1, Dalia Di Nunzio1
1Department of Clinical and Experimental Medicine, University of Messina, Messina, Italy.
Abstract:
A 58-year-old woman received a dual chamber pacemaker (Medtronic) for sick sinus syndrome. Given intact AV conduction the Managed Ventricular Pacing mode algorithm (MVP) was programmed. The day after, she suffered from palpitations. Her ECG showed a possible loss of atrial capture accompanied by atrial undersensing. Telemetry-supported pacemaker control confirmed the loss of capture. Undersensing of atrial signal was functional, related to long atrial refractory period in MVP mode algorithm. Device algorithms could induce false suspicions. Awareness about the different pacing algorithms can be useful in order to avoid erroneous interpretations and to correct potential malfunctioning.
Related Concept Videos
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Conduction System of the Heart
This system relies on the unique properties of nodal and Purkinje cells:...
Mechanism of Cardiac Arrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias