Related Experiment Video
Updated: Mar 1, 2026

Using Near-Infrared Spectroscopy Wearable Devices to Identify Central Versus Peripheral Limitations During Exercise
Published on: December 19, 2024
Validity and reliability of the Moxy oxygen monitor during incremental cycling exercise
E M Crum1, W J O'Connor1, L Van Loo2
1a School of Sport and Exercise , Massey University , Palmerston North , New Zealand.
Introduction:
The Moxy is a novel, cutaneously placed muscle oxygen monitor which claims to measure local oxygen saturation (SmO2) and total haemoglobin (THb) using near-infrared spectroscopy. If shown to be reliable, its data storage and telemetric capability will be useful for assessing localised O2 usage during field-based exercise. This study investigated the reliability of the Moxy during cycling and assessed the correlations between its measurements, whole-body O2 consumption (VO2) and heart rate (HR).
Methods:
Ten highly trained cyclists performed an incremental, step-wise cycling protocol on two occasions while wearing the Moxy. SmO2, THb, VO2 and HR were recorded in the final minute of each five-minute stage. Data were analysed using Spearman's Order-Rank Coefficient (SROC), Intraclass Correlation (ICC), and Coefficient of Variance (COV). Significance was set at p ≤ .05.
Results:
SmO2 showed a 'strong' or 'very large' correlation between trials (SROC: r = 0.842-0.993, ICC: r = 0.773-0.992, p ≤ .01) and was moderately correlated with VO2 and HR (r = -0.71-0.73, p ≤ .01). SmO2 showed a moderate to high reliability at low intensities, but this decreased as relative exercise intensity increased. THb showed poor correlations between tests and with the other measured variables, but was highly reliable at all power outputs.
Conclusions:
The Moxy is a reliable device to measure SmO2 at low to moderate intensities, but at higher intensities, greater variation in measurements occurs, likely due to tissue ischaemia or increased movement artefacts due to more frequent muscular contractions. THb has low variation during exercise, and does not appear to be a valid indicator of muscle oxygenation.
More Related Videos
09:24A Rapidly Incremented Tethered-Swimming Maximal Protocol for Cardiorespiratory Assessment of Swimmers
Published on: January 28, 2020
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Transport in the Blood