Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer
Feng Gao1, Xu Liu1, Qing Zhao1
1Department of Ophthalmology, The First People's Hospital of Xiangyang, Hubei University of Medicine, Xiangyang, Hubei 441000, P.R. China.
Abstract:
Tonometry is a fundamental procedure in routine ophthalmologic examination. Although regarded as the reference standard, the Goldmann applanation tonometer (GAT) has its limitations. A new portable alternative to the GAT is the iCare rebound tonometer (RT). The aim of the present study was to compare the intraocular pressure (IOP) results obtained using the RT and GAT and then correlate the results with the central corneal thickness (CCT). Moreover, the tolerability and safety of the RT were evaluated. The IOP of 336 patients (672 eyes) was determined by the RT and GAT. The patients were divided into three groups (group A, 7-15 mmHg, n=74; group B, 16-22 mmHg, n=218; and group C, 23-50 mmHg, n=44), based on the GAT IOP readings. Pachymetry and slit lamp inspection were also performed. To establish an agreement between the devices, a Bland-Altman analysis and paired t-test were performed. The correlation between CCT and IOP readings obtained by the two devices were assessed using linear regression correlation analysis. The mean IOP values of the RT and the GAT were 18.30±5.10 and 18.52±4.46 mmHg, respectively. There were no significant differences between them (t=-1.31, P=0.19). The 95% confidence interval of the differences between the two devices was -5.80-6.24 mmHg. The RT readings are correlated well with those of GAT (r=0.806, P=0.001). However, the RT measurements were significantly (t=-2.84, P=0.007) lower (-1.66±3.87 mmHg) than those obtained with GAT when GAT ≥23 mmHg. Both the RT (r=0.390, P=0.001) and the GAT (r=0.191, P=0.001) showed positive correlations with CCT. The IOP measurement with RT was well tolerated. None of the corneal epithelial defects was detected and all subjects denied discomfort. The RT is well tolerated and safe, and can be considered a reliable alternative to GAT for patients in a low to moderate IOP range. However, in patients with high IOP values, the measurements obtained with RT did not correlate well with those obtained by GAT. The RT readings are influenced more by CCT compared to GAT.
Related Concept Videos
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.