Uncontrolled delivery of liquid volatile anaesthetic when using the anaesthetic conserving device
Igor Karnjuš1, Dušan Mekiš2,3, Miljenko Križmarić3,4
1Department of Nursing, Faculty of Health Sciences, University of Primorska, Polje 42, 6310, Izola, Slovenia. igor.karnjus@fvz.upr.si.
Abstract:
During patient sedation with liquid volatile anaesthetic, some problems may occur through a process called auto-pumping, defined as an expansion of bubbles inside the syringe, which can lead to uncontrolled anaesthetic delivery. The study examined how the temperature of liquid volatile anaesthetics (sevoflurane and isoflurane) and the presence of gas bubbles in the syringe affect the occurrence of auto-pumping when using the anaesthetic conserving device (ACD, AnaConDa™, Sedana Medical, Uppsala, Sweden). Four different circumstances for each volatile anaesthetic were tested with a bench study: volatile anaesthetic at room temperature or precooled with and without the presence of gas bubbles in the syringe. Liquid volatile anaesthetic was infused into the ACD via a syringe pump at a fixed rate and heated gradually until the temperature of the syringe surface reached 50 °C. A main-stream gas monitor was used to measure the expired fraction of volatile anaesthetic (FE vol%). The occurrence of auto-pumping was observed only in the subgroups containing gas bubbles, with both anaesthetics. In these subgroups, the values of the expired anaesthetic gas fraction increased dramatically with the expansion of gas bubbles in the syringe (ΔFE ranged from +1.6 to 2.4 vol% for sevoflurane and +2.3 to 3.4 vol% for isoflurane). Furthermore, when the heat source was removed, a substantial decline in anaesthetic agent values below the baseline was observed with both anaesthetics. The presence of gas bubbles in the syringe, especially when exposed to a heat source, may provoke auto-pumping with uncontrolled excessive anaesthetic delivery. If auto-pumping is suspected, the syringe pump must be stopped and the ACD removed from the breathing circuit at once.
Related Concept Videos
Inhalational Anesthetics: Overview
Local Anesthetics: Clinical Application as Intravenous Regional Anesthesia
One of the advantages of...
Parenteral Anesthetics: Overview
Local Anesthetics: Clinical Application as Surface, Infiltration, and Conduction Block Anesthesia
General Anesthesia: Overview
General anesthesia induces unconsciousness in the whole body, while the others target specific areas or sensations. It is administered to minimize adverse effects, maintain...
Local Anesthetics: Clinical Application as Epidural Anesthesia
Since epidural anesthetics can be infused through an epidural catheter, all types of drugs, including short-acting ones, can be administered. Chloroprocaine and lidocaine are examples of short and long-duration anesthetics, respectively. Bupivacaine...