Full-root Aortic Valve Replacement by Stentless Aortic Xenografts in Patients with Small Aortic Roots
Reza Tavakoli1, Peiman Jamshidi2, Max Gassmann3
1Institute of Veterinary Physiology, University of Zurich; Zurich Center of Integrative Human Physiology, University of Zurich; reza.tavakoli@uzh.ch.
Abstract:
In patients with small aortic roots who need an aortic valve replacement with biological valve substitutes, the implantation of the stented pericardial valve might not meet the functional needs. The implantation of a too-small stented pericardial valve, leading to an effective orifice area indexed to a body surface area less than 0.85 cm2/m2, is regarded as prosthesis-patient mismatch (PPM). A PPM negatively affects the regression of left ventricular hypertrophy and thus the normalization of left ventricular function and the alleviation of symptoms. Persistent left ventricular hypertrophy is associated with an increased risk of arrhythmias and sudden cardiac death. In the case of predictable PPM, there are three options: 1) accept the PPM resulting from the implantation of a stented pericardial valve when comorbidities of the patient forbid the more technically demanding operative technique of implanting a larger prosthesis, 2) enlarge the aortic root to accommodate a larger stented valve substitute, or 3) implant a stentless biological valve or a homograft. Compared to classical aortic valve replacement with stented pericardial valves, the full-root implantation of stentless aortic xenografts offers the possibility of implanting a 3-4 mm larger valve in a given patient, thus allowing significant reduction in transvalvular gradients. However, a number of cardiac surgeons are reluctant to transform a classical aortic valve replacement with stented pericardial valves into the more technically challenging full-root implantation of stentless aortic xenografts. Given the potential hemodynamic advantages of stentless aortic xenografts, we have adopted full-root implantation to avoid PPM in patients with small aortic roots necessitating an aortic valve replacement. Here, we describe in detail a technique for the full-root implantation of stentless aortic xenografts, with emphasis on the management of the proximal suture line and coronary anastomoses. Limitations of this technique and alternative options are discussed.
Related Concept Videos
07:49Isolation and Analysis of Aortic Arch and Root Lesions in an Atherosclerotic Mouse Model
05:47Isolation of Mouse Interstitial Valve Cells to Study the Calcification of the Aortic Valve In Vitro
08:50Technique and Patient Selection Criteria of Right Anterior Mini-Thoracotomy for Minimal Access Aortic Valve Replacement
14:14Standardized Technique of Aortic Valve Re-implantation for Valve-sparing Aortic Root Replacement
12:17Full-root Aortic Valve Replacement by Stentless Aortic Xenografts in Patients with Small Aortic Roots
07:12Design of a Cyclic Pressure Bioreactor for the Ex Vivo Study of Aortic Heart Valves