Automatic frequency and phase alignment of in vivo J-difference-edited MR spectra by frequency domain correlation
Evita C Wiegers1, Bart W J Philips2, Arend Heerschap2
1Department of Radiology and Nuclear Medicine, Radboud university medical center, 767, PO Box 9101, 6500 HB, Nijmegen, The Netherlands. Evita.Wiegers@radboudumc.nl.
Objective:
J-difference editing is often used to select resonances of compounds with coupled spins in 1H-MR spectra. Accurate phase and frequency alignment prior to subtracting J-difference-edited MR spectra is important to avoid artefactual contributions to the edited resonance.
Materials And Methods:
In-vivo J-difference-edited MR spectra were aligned by maximizing the normalized scalar product between two spectra (i.e., the correlation over a spectral region). The performance of our correlation method was compared with alignment by spectral registration and by alignment of the highest point in two spectra. The correlation method was tested at different SNR levels and for a broad range of phase and frequency shifts.
Results:
In-vivo application of the proposed correlation method showed reduced subtraction errors and increased fit reliability in difference spectra as compared with conventional peak alignment. The correlation method and the spectral registration method generally performed equally well. However, better alignment using the correlation method was obtained for spectra with a low SNR (down to ~2) and for relatively large frequency shifts.
Conclusion:
Our correlation method for simultaneously phase and frequency alignment is able to correct both small and large phase and frequency drifts and also performs well at low SNR levels.
Related Concept Videos
2D NMR: Overview of Homonuclear Correlation Techniques
COSY90 is the standard two-dimensional (2D) COSY experiment that...
2D NMR: Overview of Heteronuclear Correlation Techniques
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are...
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
NMR Spectrometers: Resolution and Error Correction