Multicentre experience with ZSI 375 artificial urinary sphincter for the treatment of stress urinary incontinence in
Ireneusz Ostrowski1, Mariusz Blewniewski2, Frank Neugart3
1Department of Urology, Specialistic Hospital, Pulawy - Poland.
Objectives:
The aim of this study was to evaluate the safety and efficacy of the artificial urinary sphincter ZSI 375 device in male patients with stress urinary incontinence (SUI) at 10 centres.
Methods:
This was a retrospective, nonrandomised, multicentre study. From January 2012 to December 2014, and follow-up to February 2016, men with SUI were treated with the ZSI 375 device. The primary outcome was continence. The secondary outcome was complications rate.
Results:
A total of 147 patients with moderate-to-severe stress incontinence and mean age of 70.67 (26-85) years underwent implantation of the ZSI 375 device at 10 centres.The most common indication for placement was incontinence after radical prostatectomy (121 patients: 82.31%). The ZSI 375 device was considered successful in 86.40% of cases after 12 months follow-up, 83.52% after 24 months follow-up and 78.26% after 36 months follow-up. The infection rate was 6.12%, urethral erosion was 12.92% and mechanical failure was limited to 3.40% leading to ZSI 375 improvement.
Conclusions:
This study confirmed safety and efficacy of ZSI 375 device to treat moderate-to-severe SUI in men.
More Related Videos
03:55Urethroplasty with Pedicled Tunica Vaginalis for the Treatment of Long-segment Anterior Urethral Stricture Caused by Lichen Sclerosus of Glans Penis
Published on: October 18, 2024
13:45A Novel Surgical Technique in a Sheep Model for Suburethral Graft Implantation
Published on: June 20, 2025
Related Concept Videos
Urinary Tract Calculi VI: Surgical Management
Urinary Tract Calculi III: Medical Management