Related Experiment Video
Updated: Jul 31, 2026

Improving IV Insulin Administration in a Community Hospital
Published on: June 11, 2012
Intravenous Smart Pump Drug Library Compliance: A Descriptive Study of 44 Hospitals
Karen K Giuliano, Wan-Ting Su1, Daniel D Degnan2
1School of Industrial Engineering, Regenstrief Center for Healthcare Engineering, Purdue University, West Lafayette.
Background:
Although intravenous (IV) smart pumps with built-in dose-error reduction systems (DERS) can reduce IV medication administration error, most serious adverse events still occur during IV medication administration. Sources of error include overriding DERS and manually bypassing drug libraries and the DERS.
Methods:
Our purpose was to use the Regenstrief National Center for Medical Device Informatics data set to better understand IV smart pump drug library and DERS compliance. Our sample consisted of 12 months of data from 7 hospital systems, 44 individual hospitals, and descriptive data from the American Hospital Directory (AHD) for 2015. The aims of the study were (1) to determine whether there are differences in IV smart pump drug library compliance between hospital systems and (2) to provide a broad descriptive overview of relevant trends related to IV smart pump compliance.
Results:
For aim 1, we found 3 significant relationships among the 7 hospital systems: systems 3 (P < 0.001), 6 (P = 0.003), and 7 (P = 0.002) had significantly higher IV smart compliance as compared with system 4. For aim 2, the number of drug library profiles was positively correlated (P = 0.029) with IV smart pump compliance and the IV smart pump type used was significantly correlated (P = 0.013) with IV smart pump compliance.
Conclusions:
Our findings support that there are differences in IV smart pump compliance both within and between hospital systems and that IV smart pump type and the number of drug library profiles may be influencing factors. Further research is required to more accurately identify the impact of these factors in this very important area of patient safety.
Related Concept Videos
Routes of Drug Administration: Parenteral
The intravenous route (IV) of drug administration can be further categorized into two types. The bolus injection administers the entire dose rapidly, while an intravenous infusion slowly delivers smaller doses steadily.
The IV route is often...
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
Drug Accumulation During Multiple Dosing: Repetitive IV Injections
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
Parenteral Drug Delivery Systems: Injectables, Implants, and Infusion Devices