Related Experiment Video
Updated: Mar 1, 2026

Cytotoxic Efficacy of Photodynamic Therapy in Osteosarcoma Cells In Vitro
Published on: March 18, 2014
Pseudohypophosphatemia associated with high-dose liposomal amphotericin B therapy
James A Mays1, Dina N Greene1, Anne Poon2
1Department of Laboratory Medicine, University of Washington School of Medicine, Seattle, WA, United States.
Background:
Hypophosphatemia is commonly observed in critically ill patients. Inorganic phosphorus is quantified by spectrophotometric measurement of a phosphomolybdate complex, a method with multiple documented interferents. Our clinical laboratory was contacted to investigate a case of asymptomatic hypophosphatemia in a patient receiving high-dose liposomal amphotericin B therapy (L-AMB).
Methods:
In vitro experiments were performed by spiking L-AMB into residual plasma specimens. Phosphate was measured on the Beckman Coulter AU and Ortho Diagnostics Vitros instruments.
Results:
When measured on the AU, phosphate in plasma with approximately 250mcg/mL of L-AMB demonstrated a median negative bias of 3.45mg/dL relative to unspiked samples. In contrast, Vitros phosphate measurements demonstrated excellent agreement for specimens with and without L-AMB (median bias -0.2mg/dL).
Conclusions:
High L-AMB concentrations induced a significant negative bias on phosphate measured by the AU assay, but did not affect the Vitros assay. Laboratorians and clinicians should be aware of this phenomenon in patients receiving L-AMB who develop unexplained hypophosphatemia.
More Related Videos
Related Concept Videos
Chronic Kidney Disease II: Clinical Manifestations
Lysosomal Hydrolases
Nephrotic Syndrome I : Introduction
Impact of Pharmacokinetic–Pharmacodynamic Models: Regulatory Decisions
Fungal Phylum Microsporidia
Peritoneal Dialysis III: Nursing Management