Implementation of ultra-low-dose lung protocols in CT-guided lung biopsies: feasibility and safety in the clinical
Barbara K Frisch1, Karin Slebocki1, Kamal Mammadov1
11 Department of Radiology, University Hospital of Cologne, Cologne, Germany.
Abstract:
Objective To evaluate the use of ultra-low-dose computed tomography (ULDCT) for CT-guided lung biopsy versus standard-dose CT (SDCT). Methods CT-guided lung biopsies from 115 patients (50 ULDCT, 65 SDCT) were analyzed retrospectively. SDCT settings were 120 kVp with automatic mAs modulation. ULDCT settings were 80 kVp with fixed exposure (20 mAs). Two radiologists evaluated image quality (i.e., needle artifacts, lesion contouring, vessel recognition, visibility of interlobar fissures). Complications and histological results were also evaluated. Results ULDCT was considered feasible for all lung interventions, showing the same diagnostic accuracy as SDCT. Its mean total radiation dose (dose-length product) was significantly reduced to 34 mGy-cm (SDCT 426 mGy-cm). Image quality and complication rates ( P = 0.469) were consistent. Conclusions ULDCT for CT-guided lung biopsies appears safe and accurate, with a significantly reduced radiation dose. We therefore recommend routine clinical use of ULDCT for the benefit of patients and interventionalists.