Related Experiment Video
Updated: Feb 28, 2026

Esophageal Heat Transfer for Patient Temperature Control and Targeted Temperature Management
Published on: November 21, 2017
Comparison of Temporal Artery Versus Rectal Temperature in Emergency Department Patients Who Are Unable to
Carmen Brosinski1, Sherwin Valdez1, Autumn Riddell1
1Portsmouth, VA; San Diego, CA.
Introduction:
In the emergency department, pediatric and geriatric patients who present with illnesses and are unable to participate in oral evaluation of temperature must undergo a rectal temperature (RT) assessment. This study asks if a temporal artery temperature (TAT) measure can supplant the RT measure.
Methods:
A convenience sample, using a within-subject design, was used to evaluate the efficacy of TAT compared with RT in patients ≤ 3 and ≥ 65 years of age, who were unable to participate in oral temperature assessments.
Results:
Instrument reliability of the TAT is adequate for both the pediatric and geriatric populations. An unadjusted TAT did not provide acceptable temperature measurements. We also found that adjusting a TAT reading by adding -17.22°C (1° F) rendered the TAT average (either mean or median) adequately similar to RT averages for research purposes for both pediatric and geriatric groups.
Discussion:
No influence was detected on the differences between RT and TAT due to age, sex, or emergency severity index (ESI) score in patients or due to profession, years of education, or years of experience in caregivers for either the pediatric or geriatric groups. Furthermore, the adjusted TAT reading could detect fever in individual patients adequately in both the pediatric and geriatric groups. However, the adjusted TAT readings were too frequently divergent from RT readings to be used to measure temperature in individual patients for both pediatric and geriatric groups.
More Related Videos
04:34Mouse Body Temperature Measurement Using Infrared Thermometer During Passive Systemic Anaphylaxis and Food Allergy Evaluation
Published on: September 14, 2018
08:22Using an Ingestible Telemetric Temperature Pill to Assess Gastrointestinal Temperature During Exercise
Published on: October 7, 2015
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...