Related Experiment Video
Updated: Feb 28, 2026

Intra-Operative Behavioral Tasks in Awake Humans Undergoing Deep Brain Stimulation Surgery
Published on: January 6, 2011
Accuracy of Intraoperative Computed Tomography during Deep Brain Stimulation Procedures: Comparison with
Maarten Bot1, Pepijn van den Munckhof, Roy Bakay
1Department of Neurological Sciences, Rush University Medical Center, Chicago, IL, USA.
Objective:
To determine the accuracy of intraoperative computed tomography (iCT) in localizing deep brain stimulation (DBS) electrodes by comparing this modality with postoperative magnetic resonance imaging (MRI).
Background:
Optimal lead placement is a critical factor for the outcome of DBS procedures and preferably confirmed during surgery. iCT offers 3-dimensional verification of both microelectrode and lead location during DBS surgery. However, accurate electrode representation on iCT has not been extensively studied.
Methods:
DBS surgery was performed using the Leksell stereotactic G frame. Stereotactic coordinates of 52 DBS leads were determined on both iCT and postoperative MRI and compared with intended final target coordinates. The resulting absolute differences in X (medial-lateral), Y (anterior-posterior), and Z (dorsal-ventral) coordinates (ΔX, ΔY, and ΔZ) for both modalities were then used to calculate the euclidean distance.
Results:
Euclidean distances were 2.7 ± 1.1 and 2.5 ± 1.2 mm for MRI and iCT, respectively (p = 0.2).
Conclusion:
Postoperative MRI and iCT show equivalent DBS lead representation. Intraoperative localization of both microelectrode and DBS lead in stereotactic space enables direct adjustments. Verification of lead placement with postoperative MRI, considered to be the gold standard, is unnecessary.