Titanium incus interposition ossiculoplasty: audiological outcomes and extrusion rates
Selcuk Mulazimoglu1, Alexander Saxby2, Christoph Schlegel1
1Department of Otorhinolaryngology Head and Neck Surgery, Luzerner Kantonsspital, Spitalstrasse, 6000, Lucerne, Switzerland.
Summary
The Fisch titanium incus prosthesis offers reliable ossiculoplasty outcomes, with 75% of cases remaining stable long-term. This titanium prosthesis is a viable alternative to autologous incus for improving hearing.
Area of Science:
- Otolaryngology
- Biomaterials Science
- Audiology
Background:
- Ossiculoplasty aims to reconstruct the middle ear ossicular chain.
- Titanium incus prostheses offer a prosthetic alternative to autologous incus reconstruction.
- Long-term outcomes and failure rates of such prostheses require thorough evaluation.
Purpose of the Study:
- To evaluate the efficacy and safety of the Fisch titanium incus prosthesis in ossiculoplasty.
- To assess long-term audiological outcomes and identify factors influencing prosthesis success or failure.
- To compare the Fisch prosthesis with autologous incus use.
Main Methods:
- Prospective data collection from 139 patients undergoing ossiculoplasty with the Fisch titanium incus prosthesis (2001-2016).
- Analysis of 126 patients with at least 6 months follow-up (mean 4.5 years).
- Categorization into 'extrusion', 'failure' (reoperation), and 'stable' groups; audiological outcomes (air-bone gap) were measured.
Main Results:
- 75% of prostheses remained stable, 7% extruded, and 18% failed requiring reoperation.
- Mean postoperative air-bone gap (0.5-3 kHz) for the stable group was 15.3 dB.
- Atelectasis was the most common cause of extrusion; reoperations occurred at a mean of 30.7 months.
- No significant difference in outcomes based on cholesteatoma presence, surgical approach, or temporal bone pneumatization.
Conclusions:
- The Fisch titanium incus prosthesis is a reliable option for interposition ossiculoplasty, yielding hearing outcomes comparable to autologous incus.
- A stable prosthesis is associated with good long-term audiological results, with 81% achieving an air-bone gap below 20 dB.
- An expected air-bone gap of approximately 15 dB should be anticipated when using this titanium prosthesis.