Bolus Insulin Dose Error Distributions Based on Results From Two Clinical Trials Comparing Blood Glucose Monitoring
Scott Pardo1, Nancy Dunne1, David A Simmons1
11 Ascensia Diabetes Care, Parsippany, NJ.
Background:
In 2 previous clinical trials, fingertip capillary blood samples were evaluated using prespecified blood glucose monitoring systems (BGMSs) and a reference YSI glucose analyzer. In post hoc analyses, hypothetical insulin doses were calculated using these blood glucose measurements; dosing errors were compared for each trial.
Method:
For each blood glucose measurement, premeal bolus insulin dosing was determined for a hypothetical person, assuming a 60-g carbohydrate meal and 100-mg/dL target blood glucose level (adjusting 1/25 insulin sensitivity and 1/15 insulin:carbohydrate ratio inputs to account for BGMS measurement error). Dosing error was the difference between doses calculated using the BGMS and YSI results.
Results:
In Clinical Trial 1, 95% dose error ranges (in units of insulin) were: CONTOUR®NEXT EZ BGMS (EZ), -0.9 to 0.5; Accu-Chek® Aviva BGMS (ACA), -0.5 to 1.8; FreeStyle Freedom Lite® BGMS (FFL), -3.2 to -0.3; OneTouch® Ultra®2 BGMS (OTU2), -4.1 to 0.3; and Truetrack® BGMS (TT), -3.9 to 2.2. In Clinical Trial 2, these ranges were: CONTOUR®NEXT BGMS (CN), -0.7 to 1.7; Accu-Chek® Aviva Nano BGMS (ACAN), -1.3 to 1.8; FreeStyle Lite® BGMS (FSL), -5.1 to 0.2; OTU2, -1.9 to 1.2; OneTouch® Verio® Pro BGMS (OTVP), -1.0 to 1.9; and TT, -5.1 to 1.7. Within each trial, EZ and CN had statistically significantly smaller insulin dose error ranges than other BGMSs ( P <0.0001).
Conclusions:
The ranges of insulin dose errors were statistically significantly smaller with EZ and CN than with all other BGMSs in this post hoc analysis. Differences in BGMS accuracy could result in clinically important differences in insulin dosing.
More Related Videos
08:32Studying the Hypothalamic Insulin Signal to Peripheral Glucose Intolerance with a Continuous Drug Infusion System into the Mouse Brain
Published on: January 4, 2018
07:30Homogeneous Time-resolved Förster Resonance Energy Transfer-based Assay for Detection of Insulin Secretion
Published on: May 10, 2018
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Two-Compartment Open Model: IV Bolus Administration
The disparity between drug input and the sum of drug transfer rates between...
Dosage Regimen: Individualization
Errors occurring during blood pressure monitoring
Several factors...
One-Compartment Open Model for IV Bolus Administration: Estimation of Elimination Rate Constant, Half-Life and Volume of Distribution
One-Compartment Open Model for IV Bolus Administration: General Considerations
The drug's presence in the body is defined by an equation representing the difference between the rates of drug entry and exit. Key parameters—elimination rate constant,...