Related Experiment Video
Updated: Feb 28, 2026

Treating Low Back Pain in Failed Back Surgery Patients with Multicolumn-lead Spinal Cord Stimulation
Published on: June 26, 2018
Cephalad Lead Migration During a Spinal Cord Stimulation Trial: A Case Presentation
Ogoegbunam Agubuzu1, William J Beckworth2
1Department of Rehabilitation Medicine, Emory University School of Medicine, 1441 Clifton Rd, Atlanta, GA 30322.
Abstract:
Spinal cord stimulation is used in the treatment of a variety of pain conditions. Lead migration is among the most common complications associated with spinal cord stimulation. Although there have been reports of caudal lead migration, there have been no reports of significant cephalad lead migration during a spinal cord stimulation trial. Here we report what is potentially the first case of significant cephalad lead migration (from the initial T8 position to C2) during a spinal cord stimulation trial. This case demonstrates that significant lead migration is possible, and this case highlights the importance of adequately securing leads during a trial.
Level Of Evidence:
V.