In vitro evaluation of a new iterative reconstruction algorithm for dose reduction in coronary artery calcium scoring
Tobias Gassenmaier1, Thomas Allmendinger2, Andreas S Kunz1
1Department of Diagnostic and Interventional Radiology, University Hospital of Würzburg, Würzburg, Germany.
Background:
Coronary artery calcium (CAC) scoring is a widespread tool for cardiac risk assessment in asymptomatic patients and accompanying possible adverse effects, i.e. radiation exposure, should be as low as reasonably achievable.
Purpose:
To evaluate a new iterative reconstruction (IR) algorithm for dose reduction of in vitro coronary artery calcium scoring at different tube currents.
Material And Methods:
An anthropomorphic calcium scoring phantom was scanned in different configurations simulating slim, average-sized, and large patients. A standard calcium scoring protocol was performed on a third-generation dual-source CT at 120 kVp tube voltage. Reference tube current was 80 mAs as standard and stepwise reduced to 60, 40, 20, and 10 mAs. Images were reconstructed with weighted filtered back projection (wFBP) and a new version of an established IR kernel at different strength levels. Calcifications were quantified calculating Agatston and volume scores. Subjective image quality was visualized with scans of an ex vivo human heart.
Results:
In general, Agatston and volume scores remained relatively stable between 80 and 40 mAs and increased at lower tube currents, particularly in the medium and large phantom. IR reduced this effect, as both Agatston and volume scores decreased with increasing levels of IR compared to wFBP (P < 0.001). Depending on selected parameters, radiation dose could be lowered by up to 86% in the large size phantom when selecting a reference tube current of 10 mAs with resulting Agatston levels close to the reference settings.
Conclusion:
New iterative reconstruction kernels may allow for reduction in tube current for established Agatston scoring protocols and consequently for substantial reduction in radiation exposure.
More Related Videos
04:40Identifying Coronary Artery Calcification on Non-gated Computed Tomography Scans
Published on: August 28, 2018
08:02Author Spotlight: Enhanced Quantification of Cardiovascular Calcification Progression for Longitudinal Micro PET/CT Studies in Small Research Animals
Published on: November 15, 2024
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Acute Coronary Syndrome III: Diagnostic Studies